MODEL 1100CF: Thermocouple Simulator-Calibrator

The Model 1100CF is a sophisticated and versatile instrument which has been designed both to generate and to measure precise thermocouple and dc voltages.

Category:

Description

The Model 1100CF is a sophisticated and versatile instrument which has been designed both to generate and to measure precise thermocouple and dc voltages. It is equally useful on the production line and in the laboratory. Although its most frequent application is to calibrate temperature indicating devices such as digital thermometers, thermocouple recorders, etc., it actually functions in four separate and distinct ways.

As a Thermocouple Simulator. In this mode the built-in computing capability and memory enable the operator to simply select a thermocouple type, dial a temperature in °C or °F and automatically obtain the proper output voltage for the selected thermocouple type on terminals made of the proper thermocouple material.

As a Precision Digital Thermometer (Thermocouple Calibrator). In this operating mode the operator measures an external thermocouple’s temperature by adjusting the lever switches until null is reached as indicated on the front panel null meter.

As a linear DC Voltage Standard (DC Voltage Calibrator). Here the Model 1100CF generates precise dc voltages from ±1 µV to ±11.000 V.

As a linear DC Differential Voltmeter. In this mode an external dc voltage of from ±1 µV to ±11.000 V is precisely measured by adjusting the lever switches until null is reached.

The Model 1100CF may be remotely programmed in any of the above modes of operation, making it especially useful when calibrating large numbers of instruments or on the production line.

FEATURES

  • Simulates 4 Thermocouple types
  • Dial Temperature in Degrees C or F
  • Digital Conformity to NIST ±0.1°
  • Self-Contained Reference Junction
  • Dial Millivolts Directly
  • ±0.01% of Reading Accuracy
  • Optional Remote Programming